High-Quality OEM LH test Manufacturers Exporters – Novel Coronavirus (2019-nCoV) Nucleic Acid Test Kit (Multiple Fluorescence PCR) – INNOVITA
High-Quality OEM LH test Manufacturers Exporters – Novel Coronavirus (2019-nCoV) Nucleic Acid Test Kit (Multiple Fluorescence PCR) – INNOVITA Detail:
Product Detail:
Innovita® 2019-nCoV IgM/IgG Test is intended for the diagnosis and epidemiological monitoring of diseases caused by the novel coronavirus (2019-nCoV). ORF1ab and N gene of 2019-nCoV are detected qualitatively from throat swabs and alveolar lavage fluid samples collected from suspected cases of pneumonia, patients with suspected cases, others who need to diagnose.
The test results of this kit are for clinical reference only. It is recommended to conduct a comprehensive analysis of the condition based on the patient’s clinical manifestations and other laboratory tests.
Principle:
This kit uses one-step reverse transcription polymerase chain reaction (RT-PCR) detection technology to target the novel coronavirus (2019-nCoV) ORF1ab gene, N gene, and human internal reference gene sequence. Specific primers and taqman probes were designed in conserved regions.
Composition:
|
Composition |
48 tests / kit |
| Reaction Mix A | 792μL×1Tube |
| Reaction Mix B | 168μL×1Tube |
| Positive Control | 50μL×1Tube |
| Negative Control | 50μL ×1Tube |
Note: 1. Different batches of reagents should not be mixed.
2. Positive controls and positive controls need not be extracted
Test Procedure:
1. Nucleic acid extraction:
Commercial RNA extraction kits are available, magnetic bead extraction and spin column extraction are recommended for this kit.
2. Preparation the Reaction Mix:
● Take out the 2019-nCoV reaction Mix A/B and keep at room temperature until unfreezing;
● Take the corresponding portions (Reaction Mix A 16.5μL/T, Reaction Mix B 3.5μL/T) and mix, and then aliquot each PCR reaction with 20μL/ tube;
● Add 5μL of RNA template or negative control or positive control, then cover the tube cap;
● Place the reaction tube in the fluorescence PCR instrument, and set the negative / positive control and sample parameters for RT-PCR reaction according to the instrument operating instructions.
● Record Sample placement order
3.RT-PCR protocols:
Recommended Settings:
|
Cycle |
Time |
Temperature(℃) |
|
|
1 |
1 |
10min |
25 |
|
2 |
1 |
10min |
50 |
|
3 |
1 |
10min |
95 |
|
4 |
45 |
10s |
95 |
|
35s |
55 |
Product detail pictures:
Related Product Guide:
"Control the standard by the details, show the power by quality". Our organization has strived to establish a highly efficient and stable employees team and explored an effective high-quality command method for High-Quality OEM LH test Manufacturers Exporters – Novel Coronavirus (2019-nCoV) Nucleic Acid Test Kit (Multiple Fluorescence PCR) – INNOVITA , The product will supply to all over the world, such as: Sri Lanka, Kuwait, Sierra Leone, Since its foundation , the company keeps living up to the belief of "honest selling , best quality , people-orientation and benefits to customers. " We're doing everything to supply our customers with best services and best solutions . We promise that we are going to be responsible all the way to the end once our services begin.
Customer service staff and sales man are very patience and they all good at English, product's arrival is also very timely, a good supplier.